CV Step Setup

Latest updated: Mar 28, 2025
Video recording version: 8.0.0.490, compatible with BTS 8.0 software.

How to Set Up Steps for Cyclic Voltammetry (CV)

  • 1. Right-click the battery's corresponding channel, select "Single Start" to access the step configuration interface.

  • 2. Choose the cyclic voltammetry step and double-click the box for step time or voltage to enter the CV step settings.

  • 3. In the dropdown menu: "OCV" represents Open Circuit Voltage, while "Ref" stands for Reference.

  • 4. Example settings: initial voltage at 0 V vs. OCV, scan rate at 1 mV/s, vertex voltage 1 at 3.9 V vs. Ref, vertex voltage 2 at 2.6 V vs. Ref, 5 cycles, and voltage record at 3 mV.

  • 5. Select a current range exceeding the maximum scan current. Set the final voltage to 0 V vs. OCV (uncheck the box to disable if needed).

  • 6. After setup, click OK, then Start in the lower-right corner to begin testing.

  • 7. Post-test, open the data, click the CV icon to view the graph. Check the cycle count box, use the right-side buttons or input the cycle number to see each cycle's graph.

  • 8. Right-click the graph area to copy data of the displayed graph.

This operation video tutorial is applicable to the following products:
● CT/CTE-8000 Battery Testing System Series (CT-8002Q-5V100mA)

neware-battery-test-car-battery-test