Latest updated: Mar 28, 2025 Video recording version: 8.0.0.490, compatible with BTS 8.0 software.
How to Set Up Steps for Cyclic Voltammetry (CV)
1. Right-click the battery's corresponding channel, select "Single Start" to access the step configuration interface.
2. Choose the cyclic voltammetry step and double-click the box for step time or voltage to enter the CV step settings.
3. In the dropdown menu: "OCV" represents Open Circuit Voltage, while "Ref" stands for Reference.
4. Example settings: initial voltage at 0 V vs. OCV, scan rate at 1 mV/s, vertex voltage 1 at 3.9 V vs. Ref, vertex voltage 2 at 2.6 V vs. Ref, 5 cycles, and voltage record at 3 mV.
5. Select a current range exceeding the maximum scan current. Set the final voltage to 0 V vs. OCV (uncheck the box to disable if needed).
6. After setup, click OK, then Start in the lower-right corner to begin testing.
7. Post-test, open the data, click the CV icon to view the graph. Check the cycle count box, use the right-side buttons or input the cycle number to see each cycle's graph.
8. Right-click the graph area to copy data of the displayed graph.
This operation video tutorial is applicable to the following products: ● CT/CTE-8000 Battery Testing System Series (CT-8002Q-5V100mA)